Setting the sampling rate of measurement value, Setting the sampling rate of, Measurement value -39 – KEYENCE LK-G Series User Manual
Page 97

3-39
3
LK-G-M-NO3-E
3 Function Settings
Setting the Sampling Rate of Measurement Value
This section describes how to set the measurement sampling sycle. If you want the sampling
to be performed more quickly, modify it to perform faster or when the received light is extremely
dim, modify it to perform more slowly.
■
Setting procedure of the sampling rate
1
Hold down the
key, and select "
oPt on
" with the
key.
Set the common settings.
2
Press the
key, and select function "
A
" by the
key.
Set the sampling cycle.
3
Press the
key, and select
"
A-4
" with the
key.
The sampling rate is set as "500" µs in this example.
4
Press the
key to register the setting. Press the
key to return to the measurement state.
• The sampling rate will be doubled when two heads are connected and the sampling
rate is set as 20 µs.
• Set the sampling rate to other than 20 µs (other than 20 µs and 50 µs when transparent
object 2 is selected in the LK-G85/G80/G87/G82/G155/G150/G157/G152/G405/G400/
G407/G402, and other than 20 µs, 50 µs and 100 µs when transparent object 2 is
selected in the LK-G505/G500/G507/G502) when one of the items from transparent
object, transparent object 2, and multi-reflective object is selected in the
measurement mode of the head (
page 3-7).
• The sampling rate will be up to four times slower when the transparent object 2 is
selected.
• Do not use the measurement value display or received light waveform display of the
setup support software "LK-Navigator" when setting the sampling rate to 20 µs.
Function No.
Function (sampling rate)
A-0
20
20 µs
A-1
50
50 µs
A-2
100
100 µs
A-3
200
200 µs
A-4
500
500 µs
A-5
1000
1000 µs
OUT1
HI
GO
LO
HI
GO
LO
TIM
OUT2
TIM
OUT1
HI
GO
LO
HI
GO
LO
TIM
OUT2
TIM
OUT1
HI
GO
LO
HI
GO
LO
TIM
OUT2
TIM