Switching program nos, Switching program nos -9 – KEYENCE LK-G Series User Manual
Page 55

2-9
2
LK-G-M-NO2-E
2 Operations and Functions during Measurement
Switching Program Nos.
This section describes the Program No. switch function, which can easily change the operation
settings.
■
Display panel
1
Press the
key.
The program selection screen appears.
2
Select a program No. by using the
key.
3
Press the
key to register it, and return to the measurement state.
■
External input terminal
Program No. can be changed by using the P1, P2, and P3 of the expansion connector.
■
RS-232C interface
You can send the command from the external devices by using the RS-232C interface to
switch program Nos.
The method of switching the program Nos. varies depending on the setting of "Setting the
Program Switching Method" (page3-50) in the Environment settings.
Refer to "Expansion Connector" (page 4-5).
Refer to "Chapter 5 RS-232C" (page 5-RS-232C).
Function No.
Selection item
Operation method
"b-0"
"PANEL"
Panel operation/RS-232C interface
"b-1"
"
"
External terminal input
OUT1
HI
GO
LO
HI
GO
LO
TIM
OUT2
TIM
Note
E T