3 Function Settings
Setting the Scaling for Measurement (Calibration)
The displayed value for the measurement value can be calibrated as desired.
Set the values to display for the desired two points to calibrate.
Head A and head B can be calibrated for each OUT1 and OUT2.
■
Setting procedure of scaling
This section describes the setting procedure of the following calibration as an example.
1
Hold down the
key, and select "
oUt-1
" with the
key.
OUT 1 is set in this example.
Select " oUt-2 " when setting OUT2.
2
Press the
key, and select function "
b
" with the
key.
Set the scaling.
3
Press the
key, and select "
b-0
".
"Head A" is set in this example.
Select " b-1 " when setting head B.
Setting content
Setting value
Display
Point 1
Original input value 1 -1.0200
A1. nPt
Calibrated displayed
value 1
-1.0000
A1.d SP
Point 2
Original input value 2 0.9900
A2. nPt
Calibrated displayed
value 2
1.0000
A2.d SP
Display
Input
0.99
0.99
-1.02
-1.02
0.99
1.0000
-1.0000
-1.02
Display
Input