2 Operations and Functions during Measurement
Setting the Tolerance Comparator Value
This section describes how to read the tolerance comparator display, and its
setting procedure.
The Function of the Tolerance Settings
HIGH and LOW comparator tolerance values can be set. The measured value can be
displayed and output in 3 steps: when exceeding the HI comparator value (HI), when the value
drops below the LOW comparator value (LOW), and when the value is between the HI and
LOW comparator values (GO).
When the comparator output format is normal, the comparator output is as follows.
The comparator result of tolerance is outputted from the 12-pin I/O terminal block and the
expansion connector on the controller’s terminal panel.
Comparator status
Range
Display
HIGH
HI comparator value <
Measurement value
HI LED lights, and the measurement value is
displayed in red.
GO
LO comparator value
Measurement value HI
comparator value
GO LED lights, and the measurement value is
displayed in green.
LOW
Measurement value < LO
comparator value
LO LED lights, and the measurement value is
displayed in red.
Comparator
standby state
No LED lights, and the measurement value is
displayed as ------.
Refer to "Setting the Output Form of the Tolerance Comparator" (page 3-42) for the
comparator output form.
Refer to "Functions of the Input/Output Terminals" (page 4-2) for the external terminals.
ON
OFF
ON
OFF
ON
OFF
Display value
t
HI comparatator value
LO comparator value
HI output
GO output
LO output