3 Function Settings

Setting the Mounting Mode

The LK-G Series measures the displacement of the target by measuring the reflected laser
beam. When much of the surface of the target is specular, such as a mirror or glass, the
mirror-reflection component of the reflective light increases. In such a case, set the mirror
reflection setup to receive the specular reflection light.

In such a case, mount the head for mirror reflection.

The following two types of mounting mode are available.

Setting procedure of the mounting mode

1

Hold down the

key, and select "

kEAd-A

" with the

key.

"Head A" is set in this example.

2

Press the

key, and select function "

e

" with the

key.

Set the mounting mode.

3

Press the

key, and select "

e-0

" with the

key.

Select the diffuse reflection mode.

4

Press the

key to register the setting. Press the

key to return to the measurement state.

When the LK-G15/LK-G10 is connected, the setting of the mounting mode is invalid.

Function No.

Mounting

mode

Function

E-0

d FF-S

Diffuse
reflection

Sets head for diffuse reflection. Select this in normal cases.

E-1

rr-S

Mirror
reflection

Sets head for mirror reflection mode. Select this when the
measurement target is a mirror, glass or other specular
surfaces.