Setting able, Setting able -5 – KEYENCE LK-G Series User Manual
Page 63

3-5
3
LK-G-M-NO3-E
3 Function Settings
Setting ABLE
The LK-G Series measures the displacement of the target by measuring the laser beam
reflected from the surface of the target. The ABLE function automatically adjusts the light
intensity and the sensitivity to their optimum levels according to the conditions (color, luster,
and material) of the target surface.
The following two types of setting are available for ABLE.
■
Setting procedure of ABLE (AUTO)
1
Hold down the
key, and select "
kEAd-A
" with the
key.
"Head A" is set in this example.
2
Press the
key, and select the function "
A
" by the
key.
Set ABLE.
3
Press the
key, and select "
A-0
" with the
key.
Select "AUTO".
4
Press the
key to register the setting. Press the
key to return to the measurement state.
Function No.
Light intensity
adjustment
Function
A-0
AUTo
Auto
Automatically adjusts the light intensity to the optimum level.
Select this in normal cases.
A-1
AnUAL
Manual
The light intensity and sensitivity can be adjusted by limiting
the adjustment range to the desired range between 1 and 99.
Select this when the reflective rate of the target changes
rapidly and drastically, or when only detection of the target is
required.
OUT1
HI
GO
LO
HI
GO
LO
TIM
OUT2
TIM
HI
GO
LO
OUT2
TIM
OUT1
HI
GO
LO
TIM
OUT1
HI
GO
LO
HI
GO
LO
TIM
OUT2
TIM