beautypg.com

Setting the program switching method – KEYENCE LK-G Series User Manual

Page 108

background image

3-50

3

LK-G-M-NO3-E

3 Function Settings

Setting the Program Switching Method

Whether the program No. is switched from the display panel or from the external terminal input
is set.

Setting procedure of the program switching method

1

Press the

key, and select "

Env

" with the

key.

Set the environment settings.

2

Press the

key, and select function "

b

" with the

key.

Set the program switching method.

3

Press the

key, and select "

b-0

" with the

key.

The operation from the panel is set in this example.

4

Press the

key to register the setting. Press the

key to return to the measurement state.

To switch the program No. by using the command from the RS-232C, set this setting to
"Panel operation".

Function No.

Setting Selection

Function

b-0

PanEL

Panel

Panel operation/RS-232C

b-1

E T

Terminal input

External terminal input

OUT1

HI

GO

LO

HI

GO

LO

TIM

OUT2

TIM

OUT1

HI

GO

LO

HI

GO

LO

TIM

OUT2

TIM

OUT1

HI

GO

LO

HI

GO

LO

TIM

OUT2

TIM

Note