2 switching on the instrument, 3 configuration and calibration, The dhcp server is active in the network – Rice Lake PR5230 Fieldbus Transmitter User Manual
Page 93: The dhcp server is not active in the network, Indikatorbowser, Switching on the instrument, Configuration and calibration, Pr 5230 instrument manual commissioning
PR 5230 Instrument Manual
Commissioning
Sartorius
EN-93
5.2
Switching on the Instrument
The instrument can be put into operation and calibrated using a notebook/PC with the VNC program (on the
enclosed CD) and an Internet Browser.
When the supply voltage is applied to the instrument, the following information is displayed:
PR 5230
The instrument type is displayed (PR 5230)
BIOS version
Firmware version
Automatic display test
Weight display
E:Sense
Error message, if no load cells are connected (see also Chapter 15.1).
E:NoCom
Error message, if there is no communication with xBPI scale (see also Chapter 15.3).
When switching on the first time, the date and time must be set, select -[Date
&Time]:
Setup/
Clock
Date
2009-08-18
Time
11:24:53
5.3
Configuration and Calibration
There are following possibilities:
-
with VNC viewer (on the enclosed CD-ROM), see Chapter 5.3.4.
-
with an Internet Browser (‚Microsoft InternetExplorer’ or ‚Mozilla Firefox Webbrowser’), see Chapter 5.3.5.
The prerequisite is an installed and activated Java (Sun) ’applet’.
5.3.1
Connecting the Device to the Network and Finding out the IP address
The DHCP server is active in the network
An IP address is assigned to the device automatically.
The DHCP server is not active in the network
If the device is connected to a notebook/PC via a point-to-point connection, an IP address is negotiated via
function ‘AutoIP’.
This can take up to 2 minutes!
IndikatorBowser
The IP address can be found out using the ‘IndicatorBrowser’ (supplied on CD-ROM) and via the ‘host name’ of
the device (see also Chapter 5.3.3). The ‘host name’ is composed of the device name and the last 3 bytes of the
MAC ID. A label with the complete MAC ID is fitted inside the door of the instrument.
Hostname: PR5230-6B6A5E