PR 5230 Instrument Manual

Commissioning

Sartorius

EN-135


The data are saved and the instrument generates a corresponding message:

Weighingpoint/

xBPI-Scale Setup

Calibration status

complete

Net=Grs

2000 g

--

2000.01 g The weight is displayed in high-resolution (10x).

Leave with

.

Accept

ResError

Abort

5.5.8

xBPI Calibration with Automatic Weight Detection

Prerequisites:
-

The xBPI protocol has been selected (see Chapter 5.5.1).

-

The weighing point ‚xBPI-scale’ has been selected (see Chapter 5.5.4).

-

The platform configuration has been executed (see Chapter 5.5.3).

-

The setting in the following menu has been done: [Weighingpoint A-xBPI-Scale]-[Setup]:

[Configuration]-[Confirming adjust.]: [manual] (see Chapter 5.5.5).

-

The communication between instrument and platform is active.

Weighingpoint/

xBPI-Scale Setup

Calibration

Open with

.

Configuration

Select group of specification

Show device info

Weighingpoint/

xBPI-Scale Setup

Dead load

Set

Delete

Span

Adjust with user weight

Adjust with auto weight

Select with

and enter with

.

Following window appears:

Weighingpoint/

xBPI-Scale Setup

Calibration status

Load to small Calibration progress without weight.

Cal-Target

-10000.0 g The weight is displayed in high-resolution (10x).

--

-0.02 g