beautypg.com

Reply with ‘a’ and ‘b’ command, Pr 5230 instrument manual sma protocol – Rice Lake PR5230 Fieldbus Transmitter User Manual

Page 183

background image

PR 5230 Instrument Manual

SMA Protocol

Sartorius

EN-183

Reply with ‘A’ and ‘B’ Command
Reply format with ‘A' and 'B' commands (variable length):
:
The reply format and meaning are:

Start of reply from 'A'/'B' command

The field name is three characters long, left-adjusted and filled up with spaces on the
right, if necessary.

The following fields are sent:

“SMA” level/revision

(reply from ‘A’ command)

“MFG” manufacturer marking

(reply from the 1st ‘B’ command)

“MOD” product / model identification

(reply from the 2nd ‘B’ command)

“REV” software version

(reply from the 3rd ‘B’ command)

“SN_“ serial number

(reply from the 4th ‘B’ command)

“END”: This is always the last inscription field

(reply from the last ‘B’ command)

‘:’

Separator between field name and field content.

The data field contains up to 25 characters.

The SMA field contains

With the following meaning: level = (1, 2, etc.); revision = (1.0; 1.1; etc.)

End of reply from 'A'/'B' command


Example:

Command

Reply

A

SMA:1/1.0

B

MFG:Sartorius

B

MOD:PR5230

B

REV:01.01.9

B

SN_:148388723

B

END:


Note:

If the controlling computer sends another 'B' command:
B ?