beautypg.com

6 analog output (option pr 5230/06), 7 digital inputs, 8 digital outputs – Rice Lake PR5230 Fieldbus Transmitter User Manual

Page 223: 1 standard, 2 option, 9 serial interfaces, 1 rs-232, 2 rs-485, Analog output (option pr 5230/06), Digital inputs

background image

PR 5230 Instrument Manual

Specifications

Sartorius

EN-223

16.5.6

Analog Output (Option PR 5230/06)

Output

gross or net or fixed preset value

Range

0/4 mA…20 mA

galvanically isolated

Load

0…500 ¥

Resolution

1 µA

Linearity error

<0,05 %

Total error

<0,1 %

Combined error - temperature

<100 ppm/K

16.5.7

Digital Inputs

3 inputs galvanically isolated via optocoupler (passive mode, external supply). Can be switched to active in the
instrument, for connection of switches. No common potential in passive mode.

Input signal for 'Low'

0…5 V or open input

Input signal for 'High'

10…28 V

Input current for 'High'

<11 mA @ 24 V, <5 mA @12 V

16.5.8

Digital Outputs

16.5.8.1

Standard

3 outputs, relays

Switch-over contact

Contact rating max.

250 V AC

5,0 A

1250 VA

250 V DC

0,3 A

100 V DC

0,5 A

50 V DC

1,5 A

30 V DC

5,0 A

16.5.8.2

Option

3 Opto-decoupled outputs

Supply voltage

24 V +10 %

Output current

Max. 40 mA

Voltage drop

3,2 V @ 40 mA

16.5.9

Serial Interfaces

16.5.9.1

RS-232

Transfer rate

Max. 19200 bd

Not galvanically isolated

16.5.9.2

RS-485

Übertragungsrate

Max. 19200 bd

Galvanically isolated