4 effect of ambient conditions, 1 environmental conditions, 2 electromagnetic compatibility (emc) – Rice Lake PR5230 Fieldbus Transmitter User Manual
Page 221: 3 rf interference suppression, Effect of ambient conditions, Environmental conditions, Electromagnetic compatibility (emc), Rf interference suppression, Pr 5230 instrument manual specifications
![background image](/manuals/639397/221/background.png)
PR 5230 Instrument Manual
Specifications
Sartorius
EN-221
16.4
Effect of Ambient Conditions
16.4.1
Environmental Conditions
Temperature range
Ambient temperature for operation
-10...+40 °C
Ambient temperature 'not legal for trade'
-10...+50 °C
Power-on temperature
0...+40 °C
Storage/transport
-20...+70 °C
Humidity
<95 %, no condensation, (acc. to IEC 68-2)
Protection type
IP 66 housing
Height
<2000 m
16.4.2
Electromagnetic Compatibility (EMC)
All data in compliance with NAMUR NE 21, EN 45501 and EN 61326.
Housing
Radio frequency electromagnetic fields
(80 – 3000 MHz)
EN 61000-4-3
10 V/m
Electrostatic discharge (ESD)
EN 61000-4-2
6/8 kV
Signal and control
lines
Fast transients (burst)
EN 61000-4-4
3 kV
Peak voltages (surge) 1.2/50 µs
EN 61000-4-5
2 kV
Conducted disturbances by radio frequency
(0.15 – 80 MHz)
EN 61000-4-6
10 V
Mains inputs
Fast transients (burst)
EN 61000-4-4
3 kV
Peak voltages (surge) 1.2/50 µs
EN 61000-4-5
1/2 kV
Conducted disturbances by radio frequency
(0.15 – 80 MHz)
EN 61000-4-6
10 V
Voltage dips
EN 61000-4-11
0/40/70 %
20/200/500 ms
Short interruptions
EN 61000-4-11
5 s
16.4.3
RF Interference Suppression
Electromagnetic emission
In acc. with EN 61326, limit value class A