11 determination test value/display test value, 12 finishing/saving the calibration, Determination test value/display test value – Rice Lake PR5230 Fieldbus Transmitter User Manual
Page 121: Finishing/saving the calibration, 12, in, Pr 5230 instrument manual commissioning
PR 5230 Instrument Manual
Commissioning
Sartorius
EN-121
5.4.11
Determination Test Value/Display Test Value
Press [Calc Test] activates test value calculation. The maximum capacity (Max) is displayed with the
designation
without a weight unit.
The value determined during calibration after starting the test [CalcTest] is shown.
Weighingpoint/WP A/
Calibration
Calc Test
Depending on the settings under
-[Weighingpoint]-[Calib]-[Param]-[Test mode], either
-
the maximum capacity with [Absolute], or
-
the deviation from the test value with [Relative]
is shown when you press
to view the test data.
5.4.12
Finishing/Saving the Calibration
Finish the calibration with
.
You are prompted to confirm whether calibration should be closed without
determining the test value.
If not all data was determined when calibrating with [New] (e.g. dead load
not set/entered), this message is shown:
Press [Yes] to confirm and then press again; another prompt is displayed:
Press [Save] to save changes in calibration data. If you press [Undo], changes
are not saved and the display returns to the selection menu for the
weighing points.
The confirmation is displayed as follows:
Saving calibration
After quitting the menu, the following message is displayed:
Exit calibration
After finishing the calibration, set the CAL switch to the closed position; see also Chapter 5.1.1.