2 weight error status, 3 error messages with xbpi scales, Weight error status – Rice Lake PR5230 Fieldbus Transmitter User Manual
Page 215: Error messages with xbpi scales, Er 15.3), Pr 5230 instrument manual error messages
PR 5230 Instrument Manual
Error Messages
Sartorius
EN-215
15.2
Weight Error Status
VNC text (see also Chapter 15.1)
E:Arith
E:Invrs
E:Overl
E:NoSig
E:Sense
E:HardE/ E:NoCom
flashing 1 Hz flashing 1 Hz
Altern. flash. 1 Hz Altern. flash. 1 Hz
flashing 1 Hz
flashing 1 Hz
Altern. flash. 1 Hz
flashing 1 Hz flashing 1 Hz flashing 1 Hz flashing 1 Hz
Altern. flash. 1 Hz Altern. flash. 1 Hz
Note:
With all other messages, the upper status indicator LED blinks.
15.3
Error Messages with xBPI Scales
Indication
VNC text
Error and possible cause
Remote display
E:Value
Value exceeds display The weight value is not displayed:
-
Too many digits have been set
Error 4
E:NoCom
No values from scale No Communication with xBPI scale:
-
Cable break
-
Internal scale error
-
The scale is not connected to the supply
voltage.
Error 9
E:NoWgt
No weight data
No weight data:
-
Wägepunkt umgeschaltet
• Mit
den neuen Wägepunkt dem Gerät
zuweisen.
Error •
E:NotRed
Scale not ready
The scale is not ready for weighing:
-
The instrument is in the warm-up phase.
-
The instrument is in the automatic taring
mode.
-
The instrument was switched on with the
scale loaded.
• Switch the device off and on again.
Error c
E:BadDev
Wrong serial number ‘Serial number’ does not correspond to the
number set in the device.
Error ?