PR 5230 Instrument Manual

Error Messages

Sartorius

EN-215

15.2

Weight Error Status

VNC text (see also Chapter 15.1)

E:Arith

E:Invrs

E:Overl

E:NoSig

E:Sense

E:HardE/ E:NoCom

flashing 1 Hz flashing 1 Hz

Altern. flash. 1 Hz Altern. flash. 1 Hz

flashing 1 Hz

flashing 1 Hz

Altern. flash. 1 Hz

flashing 1 Hz flashing 1 Hz flashing 1 Hz flashing 1 Hz

Altern. flash. 1 Hz Altern. flash. 1 Hz


Note:

With all other messages, the upper status indicator LED blinks.

15.3

Error Messages with xBPI Scales

Indication

VNC text

Error and possible cause

Remote display

E:Value

Value exceeds display The weight value is not displayed:

-

Too many digits have been set

Error 4

E:NoCom

No values from scale No Communication with xBPI scale:

-

Cable break

-

Internal scale error

-

The scale is not connected to the supply

voltage.

Error 9

E:NoWgt

No weight data

No weight data:
-

Wägepunkt umgeschaltet

• Mit

den neuen Wägepunkt dem Gerät

zuweisen.

Error •

E:NotRed

Scale not ready

The scale is not ready for weighing:
-

The instrument is in the warm-up phase.

-

The instrument is in the automatic taring

mode.

-

The instrument was switched on with the

scale loaded.

• Switch the device off and on again.

Error c

E:BadDev

Wrong serial number ‘Serial number’ does not correspond to the

number set in the device.

Error ?