PR 5230 Instrument Manual
Commissioning
Sartorius
EN-127
5.5.2
xBPI Scale Function
Weighingpoint
Weighingpoint A
xBPI-Scale Select [xBPI-Scale] with
and
.
Select [Config].
Setup
Config
Param
The menu appears.
Weighingpoint/
WP A
Type
xBPI-Scale
W&M
none Select the W
&
M mode with
.
Tare timeout
1.0 s
Serial number
0
SBN Address
0
Leave with
.
[Tare timeout]
Waiting time for the execution of a zero set or taring command. If the xBPI-scale has not executed the
command in the specified time, the action will be aborted.
[Serial number]
Serial number of the connected xBPI scale/weighing module. The number is required for checking with legal
for trade application. With serial number 0, checking is omitted.
[SBN Address]
With an address unequal to 0, bus operation active, possible addresses: 1 - 31, i.e. max. 31 xBPI scales can be
operated at an RS 485 bus line. The SBN Address is shown in the display.
Example: Address 31 at WP-A.
5.5.3
xBPI Platform Configuration
Weighingpoint
Weighingpoint A
xBPI-Scale Select [xBPI-Scale] with
and
.
Select [Setup].
Setup
Config
Param
Read the parameters from the xBPI scale with [Setup].
Weighingpoint/
xBPI-Scale Setup
Reading parameters
model
metrologie
Ticks indicate the progress.
device info
settings
An error message displays, unless communication with the xBPI scale is possible!