Commissioning

PR 5230 Instrument Manual

EN-124

Sartorius

5.4.13.1

Legal-for-Trade Operation

Under

-[Weighingpoint]- [Calib]-[Param]-[W

&M] you can choose between:

[none] a legal-for-trade mode [OIML], [NTEP] or [NSC].

[none]

[OIML]

[NTEP]

[NSC]

Gross weight display

B

B

G

G

Min. meas. signal

0,125 mV/V at
30000 d

0,125 mV/V at
3000 e

0,125 mV/V at
3000 e

0,125 mV/V at
3000 e

0,25 mV/V at
60000 d

0,25 mV/V at
6000 e

0,25 mV/V at
6000 e

0,25 mV/V at
6000 e

0,42 mV/V at
10000 e

0,42 mV/V at
10000 e

0,42 mV/V at
10000 e

If legal-for-trade operation is selected, the parameters (zero tracking etc.) must be set accordingly; they are
not checked.
The relevant CAL switches (see Chapter 5.1.1) must be sealed in the closed position.
Note:

If the multi-interval scale [Multi-interval] has been selected in the menu [Range mode], the
W&M mode [OIML] cannot be selected.
With [OIML] selection, the following message is displayed:

W&M mode OIML not

allowed as long as Range

mode is multi interval


When used in legal metrology, no more than 3 decimals are permissible.

If a max. value with more than 3 decimals has been entered during
calibration, the following message is displayed.
Moreover, the following message is shown, if [OIML] has been selected:

W&M mode OIML not

allowed as with max of

more than 3 decimals.


Note:

In W&M mode an invalid weight is shown without mass unit.