beautypg.com

2 leds on the fielbus card, Leds, Legend – Rice Lake PR5230 Fieldbus Transmitter User Manual

Page 87: 3 connection

background image

PR 5230 Instrument Manual

Installing the Instrument and Plug-in Cards

Sartorius

EN-87

4.4.8.2

LEDs on the fielbus card

Watchdog LED

Meaning

Frequency

Color

Flashing

1 Hz green

Module is initialized and ready for operation.

Flashing

2 Hz green

Module is not initialized.

Flashing

1 Hz red

Check error in ASIC and FLASH ROM, module is defect.

LEDs

1 2

3 4

LED 1

LED 2

LED 3

LED 4

Lit

- - -:

No link or power off.

Lit

- - -:

Offline, no connection.

Lit

- - -:

No function.

Lit

- - -:

No power or not initialized.

Lit

grn:

Link established.

Lit

grn:

Link established and
online connected.

Lit

grn:

Initialized, no error.

Flashing 1 Hz

grn:

Receiving/Transmitting data.

Flashing 1 Hz

grn:

Online, no connection.

Flashing 1 Hz

grn:

Diagnostic data available.

Flashing 2 Hz

grn:

Engineering tool to identify
is active.

Flashing 1 Hz

red:

Configuration error.

Flashing 3 Hz

red:

No station name,
no IP address

Flashing 4 Hz

red:

Internal error.

Legend

---

off

red

red

grn

green

4.4.8.3

Connection

• Take the cable (e.g. patch cable CAT5) into the instrument through

the metal sleeve of the cable gland, strip the insulation and mount
the delivered RJ-45 plug (see mounting informations).

• Put the RJ-45 plug into the RJ-45 connecting socket of the fieldbus

card.

• Tighten the cable gland of the instrument.