Pr 5230 instrument manual commissioning – Rice Lake PR5230 Fieldbus Transmitter User Manual

Page 137

background image

PR 5230 Instrument Manual

Commissioning

Sartorius

EN-137


Following window appears:

Weighingpoint/

xBPI-Scale Setup

Calibration status

Load to small

Cal-Target

-5000.0 g

--

0.00 g

The following window is displayed after applying the weight:

Weighingpoint/

xBPI-Scale Setup

Calibration status

Difference display Calibration progress without weight.

Cal-Delta

-0.0 g

--

4999.98 g The weight is displayed in high-resolution (10x).

Select [Accept].

Accept

ResError

Abort

The data are saved and the instrument generates a corresponding message:

Weighingpoint/

xBPI-Scale Setup

Calibration status

complete

Net=Grs

5000.0 g

--

5000.00 g The weight is displayed in high-resolution (10x).

Leave with

.

Accept

ResError

Abort