Commissioning
PR 5230 Instrument Manual
EN-106
Sartorius
5.3.8.1
Weighingpoint ‚Internal A’
- Calibration
Calibration of weighing electronics
- New
Reset Span and dead load
Contin, Cancel
- Max
0.00001 ...<3000>... 999999
- Scale interval
<1>, 2, 5, 10, 20, 50
- Dead load at
<0.000000 mV/V> or [by load]
- Max at
<1.000000mV/V> or [by load]
by load
0.00001 ... 999999
- Calibrated at
(Display only)
- Sensitivity (µV/d)
(Display only)
- Test
Determine test value
Exit calibration
Save or cancel changes
- Modify
Only for minor modifications/ setting new dead load, otherwise [New]
see New
- Param
- Measuretime
5 ms, 10, 20, 40, 80, 160, <320>, 640, 960, 1280, 1600ms
- Digital filter
- External supply
<8 -12 VDC>, below 8 VDC
- * Fcut
Cut off frequency, *only unless filter 'off', 0.1 - 80.0 Hz
- Test mode
- W
& M
been selected or Max has more than 3 decimals), NSC, NTEP
- Standstill time
0.01 s…<0.50 s> ... 2.0 s (range is dependent on response time)
- Standstill range
0.00 d ... <1.00 d> ... 10.00 d
- Tare timeout
0.1 s ... <2.5 s> ... 25 s, timeout due to instability
- Zeroset range
0.00 d ... <50.00 d> ... 10000.00 d
- Zerotrack range
0.00 d ... <0.25 d> ... 10000.00 d
- Zerotrack step
0.00 d ... <0.25 d> ... 10.00 d
- Zerotrack time
<0.0 s> ... 25 s
- Overload
0 d ... <9 d> ... 999999 d
- Min
0 d ... <50 d> ... 999999 d, minimum weight
- Range mode
- * Range limit 1
In weight, unit same as Max, transition from small to medium scale
interval
- * Range limit 2
In weight, unit same as Max, transition from medium to high scale
interval
* only for multiple range or multi-interval
- View(when CAL switch closed)
- Max
(Display only)
- Scale interval
(Display only)
- Dead load at
(Display only)
- Max at
(Display only)
- Calibrated at
(Display only)
- Sensitivity (µV/d)
(Display only)
- Param
Items as for Param. (display only)