Pr 5230 instrument manual commissioning – Rice Lake PR5230 Fieldbus Transmitter User Manual

Page 155

background image

PR 5230 Instrument Manual

Commissioning

Sartorius

EN-155

Function

SPM bit

Description

actual diginp1

X00=1

Digital input 1: active

actual diginp2

X01=1

Digital input 2: active

actual diginp3

X02=1

Digital input 3: active

actual limit 1

X16=1

Limit signal 1: active

actual limit 2

X17=1

Limit signal 2: active

actual limit 3

X18=1

Limit signal 3: active

ADC error

X32=1

General error in the weighing point

above Max

X33=1

Weight above Max

overload

X34=1

Weight above Max plus the ‘overload’ value

below zero

X35=1

Weight below zero

center zero

X36=1

Weight within ¼ d of zero

inside ZSR

X37=1

Weight within zero-setting range

standstill

X38=1

Mechanical stability of the scale

out

X39=1

Weight below zero or above Max

command error

X48=1

For internal use only.

command busy

X49=1

For internal use only.

power fail

X50=1

Set after power-on (=power failure)

test active

X56=1

Analog test was started.

cal active

X57=1

For internal use only.

tare active

X58=1

Instrument is tared.

marker bit 1

X64=1

Marker bit 1 set, after power-on the markers are set to ‘0’.

marker bit 2

X65=1

Marker bit 2 set, after power-on the markers are set to ‘0’.

marker bit 3

X66=1

Marker bit 3 set, after power-on the markers are set to ‘0’.

Example: ‚overload’

X34=1

Function and output are active
(e. g.: if ‘overload’ is reached, a lamp is lit).

X34=0

Function and output are not active
(e. g.: if ‘overload’ is reached, a lamp is lit).