beautypg.com

5 communication error, Communication error, Pr 5230 instrument manual sma protocol – Rice Lake PR5230 Fieldbus Transmitter User Manual

Page 185

background image

PR 5230 Instrument Manual

SMA Protocol

Sartorius

EN-185


Example: 6000 kg x 1 kg platform scale

Command

Reply

I

SMA:2/1.0

N

TYP:S

N

CAP:kg_:6000:1:0

N

CMD:HPTMCR

N

END:


Example: 5000g x 1 g, 10000g x 2, 25000 g x 5 multiple range/multi-interval

Command

Reply

I

SMA:2/1.0

N

TYP:S

N

CAP:g__:5000:1:0

CAP:g__:10000:2:0

CAP:g__:25000:5:0

N

CMD:HPTMCRQ

N

END:

7.5

Communication Error

When a communication error due to a parity error or a frame error (if used) was detected, the scale sends an
ASCII ‘!’ character. The only other error is the detection of an unknown or non-supported command. In this
case, an ASCII '?' character is sent as a reply. Dependent on the error messages, the controlling computer must
decide how to continue the scale operation.