beautypg.com

5 commissioning, 1 data protection/power failure, 1 cal switch – Rice Lake PR5230 Fieldbus Transmitter User Manual

Page 92: Commissioning, Data protection/power failure, Cal switch, 5commissioning, Pr 5230 instrument manual

background image

Commissioning

PR 5230 Instrument Manual

EN-92

Sartorius

5

Commissioning

The control and display assignment is described in Chapter 2.3.

5.1

Data Protection/Power Failure

The calibration data and parameters as well as all configuration and interface data are stored in non-volatile
(EAROM) memory. Unauthorized data changing can be prevented by an access code. Additional write
protection is provided for calibration data and parameters (CAL switch, see Chapter 5.1.1).
Clock and calendar continue running in the event of a power failure.

5.1.1

CAL Switch

The CAL switch is located on the weighing electronics board. Generally, we recommend setting the switch into
the closed position after calibration to prevent accidental overwriting/data loss.
With legal-for-trade applications, the CAL switch must be sealed in the closed position.

Note:

If the weighing electronics board has been changed after calibration or the device is not calibrated
the weight value display will show ‘E:BadDev’ if the CAL switch is closed.

Status

Indication

Switch settings

[opened]

Weight value normal

[closed]

Weight value invers

To view the position of the CAL switch, select -[Show status]:

Info/

Status

Free system RAM

5480 of 15204 kb

Cal-Switch

opened [opened] = opened; no write protection

Builtin RS232

no signal [closed] = closed; write protection is active