Register 2: state of state-controlled action bits, Register 4: calibration information, error byte, Pr 5230 instrument manual fieldbus interface – Rice Lake PR5230 Fieldbus Transmitter User Manual
Page 197
![background image](/manuals/639397/197/background.png)
PR 5230 Instrument Manual
Fieldbus Interface
Sartorius
EN-197
Register 2: State of State-Controlled Action Bits
Only reading is permitted; the signal state is shown.
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
Byte 1
Byte 2 87:
GetFixTare
86:
SetFixTare
85:
ResetPWF
84:
ResetTest
83:
SetTest
82:
ResetTare
81:
SetTare
80:
SetZero
Byte 3
89:
ResetError
88:
Print
Register 4: Calibration Information, Error Byte
Only reading is permitted.
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0 EXPO
Byte 1 UNIT
Byte 2 STEP
Byte 3 LASTERROR
EXPO
One byte for the position of the decimal point; content in decimal form 0 ... 255
0 = 000000
1 = 00000.0
2 = 0000.00
3 = 000.000
4 = 00.0000
5 = 0.00000
UNIT
One byte for the weight unit; content in decimal form: 0 ... 255
2 = g, grams 3 = kg, kilograms
4 = t, tons 5 = lb, pound
STEP
One byte for scale interval; content in decimal form: 0 ...255
1 = scale interval '1', 2 = scale interval '2', 5 = scale interval '5'
10 = scale interval '10', 20 = scale interval '20', 50 = scale interval '50'
LASTERROR
LastError byte; see also
CmdError
bit, number of
'last error'
:
31 = no stability was reached (e.g. when taring)
33 = negative weight value when taring and ’legal-for-trade’ mode on
47 = no zero setting; weight not within zero setting range
107 = no stability with Getfixtare