PR 5230 Instrument Manual

Error Messages

Sartorius

EN-217

The dead load entered in mV/V plus maximum capacity in mV/V is higher
than 3 mV/V (= 36 mV).

The scale is not stable.
Remedial action
• Check the mechanical function of the scale.
• Adapt the filter setting; reduce the resolution.
• Adapt the stability conditions.

Measurement signal is negative (load cells connected with wrong polarity
or defective) when determining the dead load with [by load].
Cause
Load cell connected with wrong polarity, or defective.

The dead load entered in mV/V is higher than 5 mV/V.

The scale is not stable.
Remedial action
• Check the mechanical function of the scale
• Adapt the filter setting; reduce the resolution.
• Adapt the stability conditions.

The weight on the scale is less than the dead load after input of the weight
value.

The maximum capacity is not an integer multiple of the scale interval.