PR 5230 Instrument Manual

Installing the Instrument and Plug-in Cards

Sartorius

EN-45

4.2.7

Opto-decoupled Outputs (optional)

This option can be used within Ex areas Zone 2 and Zone 22, see also Chapter 3.1.
The main board is provided with 3 opto-decoupled outputs.
The digital outputs are passive. An external power supply is required.

-

CH1

+

-

CH2

+

-

CH3

+

4

5

3

2

1

6

Number of outputs 3 (CH 1, CH 2, CH 3 )
Output signal

Max. 24 V +10 %; external supply

Output current

Max. 40 mA

Voltage drop

3.2 V @ I

max

Connection

Plug-in 6-pin screw terminal block,
conductor cross-section max. 1.5 mm²

Cable

Shielded, max. 50 m.
The cable screen must be connected in the
sleeve of the cable gland!

Component layout on the main board
In this option the jumper R320 …R325 are

not fitted on the main board.

R321

R323

R325

R324

R322

R320