Commissioning

PR 5230 Instrument Manual

EN-136

Sartorius


In this example, a weight of 5000 g is put onto the scale.
The following window is displayed after applying the weight:

Weighingpoint/

xBPI-Scale Setup

Calibration status

Difference display

Cal-Delta

0.1 g

--

5000.06 g The weight is displayed in high-resolution (10x).

Select [Accept].

Accept

ResError

Abort

The data are saved and the instrument generates a corresponding message:

Weighingpoint/

xBPI-Scale Setup

Calibration status

complete

Net=Grs

5000.0 g

--

5000.01 g The weight is displayed in high-resolution (10x).

Leave with

.

Accept

ResError

Abort

5.5.9

xBPI Calibration with Default Weight

Prerequisites:
-

The xBPI protocol has been selected (see Chapter 5.5.1).

-

The weighing point ‚xBPI-scale’ has been selected (see Chapter 5.5.4).

-

The platform configuration has been executed (see Chapter 5.5.3).

-

The setting in the following menu has been done: [Weighingpoint A-xBPI-Scale]-[Setup]:

[Configuration]-[Confirming adjust.]: [manual] (see Chapter 5.5.5).

-

The communication between instrument and platform is active.

Weighingpoint/

xBPI-Scale Setup

Calibration

Open with

.

Configuration

Select group of specification

Show device info

Weighingpoint/

xBPI-Scale Setup

Adjust with default weight

Select with

and enter with

.

Adjust with intern weight

Linearity

Default

User