beautypg.com

Commissioning, Pr 5230 instrument manual – Rice Lake PR5230 Fieldbus Transmitter User Manual

Page 166

background image

Commissioning

PR 5230 Instrument Manual

EN-166

Sartorius


Following window appears:

4.

Click the file that must be loaded.

5.

Click on ‘Open’.


The file is displayed in the window.


6.

Click on ‚Restore’.

The selected file is loaded into the device.
7.

Press the softkey ‚Modify‘ in the menu [Setup]-[Weighingpoint]-[Internal]-[Calib.].

8.

Change a value (e.g. scale interval).

9.

Press ‘Exit’ several times until you are prompted to confirm whether the changes should be saved.

10.

Press the the softkey ‚Yes‘. When saving, the copied serial number is overwritten with the serial number of

the weighing electronics board provided in this instrument.

11.

Undo the change and save it.

12.

Close the CAL switch.