beautypg.com

5 load cells, 6 connection cable, 7 environmental conditions – Rice Lake PR5230 Fieldbus Transmitter User Manual

Page 37: Pr 5230 instrument manual options

background image

PR 5230 Instrument Manual

Options

Sartorius

EN-37

3.2.10.5

Load Cells

Supply voltage

±6 V DC (nominal)/

≤96 mA

Max. 4 load cells of 650

Ω each, connected in

parallel (= 162.5 ¥)

±3.6 V DC (nominal)/

≤96 mA Max. 8 load cells of 650 Ω each, connected in

parallel (= 81.3 ¥)
or
4 load cells of 350

Ω each, connected in parallel

(= 87.5 ¥)

Tolerance range

±5.8…±6.2 V DC; ±3.4…±3.8 V DC

Max. current

≤100 mA short circuit

Measuring voltage V

meas

Max. 36 mV (differential)

Sense voltage V

sense

±6 V nominal (with ref. to ground)

±3.6 V nominal (with ref. to ground)

3.2.10.6

Connection Cable

Length of the connection cable between junction box and PR 5230

Type

PR 6135, PR 6135A

Max. 200 m – length of the load cell cable

Length of the connection cable between weighing platform and PR 5230

Type

LiYCY

Max. 200 m

3.2.10.7

Environmental Conditions

Use the transmitter only within the temperature range of -10 °C…+40 °C. Avoid the inadmissible exposure of
heat, cold, direct sunlight, UV radiation or vibration. Install the instrument in that way, that air circulation is
possible and heat sources are sufficient far away.