MODEL6300
7. 12
MODEL6300
Section 7 Integration value measurement
Section
7
 
7-5 Saving data 
When integration or demand measurement starts, the measured data will be 
saved automatically. 
There are two locations where data can be saved. 
* CF card
: Max. 20 files can be saved.
* Internal memory : Only 1 file can be saved.
Data is saved to a CF card automatically when a CF card has been inserted 
before turning on the instrument. If the CF card has not been inserted, data is 
saved automatically to internal memory. 
 
7-5-1 Saving Procedure 
* When a survey is started (manually or automatically) a file is opened. 
* Data is saved at the end of each integration interval (“Setting 09”). 
Elapsed time
Integration interval
1 2 3
n-1
n
Start of survey
Data saving point
End of survey
* When the survey is closed (manually or automatically) the file is closed 
* All the recorded parameters at each data saving point are saved to one file. 
NOTE
* Never set the Function switch to OFF position during a survey otherwise the 
 survey may be lost. 
* The file number becomes “001” when; 
 (1) when the file number has exceeded 999 
 (2) after system reset 
 
* Data can also be saved to the internal memory even though a card has been 
 inserted. Refer to “Setting 19” of Section 4 in this manual. 
 
 