2 instantaneous value measurement – KYORITSU 6300 Instruction Manual User Manual
Page 142

13. 3
MODEL6300
Section 13 Specification
MODEL6300
13-2 Instantaneous value measurement
(1) Voltage Vi [V]
Range
150/ 300/ 600V
Displayed digit
4 digits
Allowable input
10 ~ 110% of each range
Display range
5 ~ 120% of each range
Crest factor
2.5 or less
Accuracy
±0.3%rdg±0.2%f.s. (sine wave 45 ~ 65Hz)
Instantaneous overload
720V (for 10 sec.)
Input impedance
Approx. 2.7M
ȍ
(2) Current Ai [A]
Range
50A type
:5/ 10/ 20/ 50A
( 500mV / 50A = 10mV / A )
100A type :10/ 20/ 50/ 100A
( 500mV / 100A = 5mV / A )
200A type :20/ 50/ 100/ 200A
( 500mV / 200A = 2.5mV / A )
500A type :50/ 100/ 200/ 500A
( 500mV / 500A = 1mV / A )
1000A type :100/ 200/ 500/ 1000A
( 500mV / 1000A = 0.5mV / A )
3000A type :1000/ 3000A (3000A: CT=3.00)
( 500mV / 1000A,3000A = 0.5mV, 0.167mV / A )
Displayed digit
4 digits
Allowable input
10 ~ 110% of each range
Display range
1 ~ 120% of each range
Crest factor
3.0 or less (each range: 90% or less)
Accuracy
±0.3%rdg±0.2%f.s. + accuracy of clamp sensor
(sine wave: 45 ~ 65Hz)
Instantaneous overload
120% of the rated current of each clamp sensor
(for 10 sec.)
Input impedance
Approx. 100k
ȍ
(3) Active power Pi [W]
Range
(Voltage range) x (Current range)
Displayed digit
4 digits
Accuracy
±0.5%rdg±0.2%f.s. + accuracy of clamp sensor
(power factor: 1, sine wave: 45 ~ 65Hz)
Effect of power factor
±1.0%rdg
(indicated value: power factor 0.5 against power factor 1)
Polarity indication
Consumption : +( no mark), Regeneration: -
(4) Frequency f [Hz]
Measuring range
40.0 ~ 70.0Hz
Displayed digit
3 digits
Allowable input
10 ~ 110% of each voltage range
(sine wave 45 ~ 65Hz)
Accuracy
r3dgt
Input source
Detect from V1 (voltage between voltage input
terminals: VN and V1)