NOTE
* The start time and date should be set after the current time in such a way to 
 give enough time to the user to complete all settings before the survey 
 starts. 
 
* When the start time and date are set before the current time, measurement 
 starts immediately upon pressing the 
START/STOP
key.
 
* When the preset start time and date come after the preset stop time and 
 date, measurement stops right after the start of measurement. 
 
* Even if the start and stop time have been preset and the instrument is in 
 standby mode, pressing the 
START/STOP
key at least 2 sec. will release
 the stand-by mode and start a survey in Manual mode. This renders the 
 start/ stop time and date settings ineffective. 
 
Then , (or ) marks disappear and the LED status 
indicator goes off. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
8. 11
MODEL6300
Section 8 Demand value measurement
MODEL6300