STREAM CONFIGURATION

06

Sensors

The number of sensors connected to the SUMMIT

Keypad

The default value which can be used in the selected

Max, Min

The alarm limits for the pressure. If only one of the two must be used, see advanced.

Hi, Lo

The warning limits for the pressure. If only one of the two must be used, see advanced.

Select

The 6 choices for selection of the pressure source

6 .6 .1 Sensor calibration constants

The calibration range and offset are the scaling factors for the transmitter, determined during

the calibration of the transmitter. As this is a multiplier to the normal transmitter range and

offset, this is not a mandatory setting, but a software calibration. This is used to correct the

transmitter value without actually calibrating the transmitter itself. The advantage is that the

changes will appear in the audit log.

This setting is also ideal for simulation purposes. With it, the transmitter can be set to any value

without an actual transmitter connected.

Figure 92 Stream pressure calibration constants

Range

Default value is 1. This is a multiplier to the actual transmitter range.

Offset

Default value is 0. This is an offset to the actual transmitter range.

6 .6 .2 Advanced

Figure 93 Stream pressure advanced options