Figure 141 Prover alarm settings, re-prove

For both the re-prove and the start of prove the following parameters can be set for volume,

pressure, temperature and density.

Limit

The % of the set value not to be exceeded for the prove to start

Duration

The time with which the limit must be reached

If these conditions are not met, the prove will not start, a report will be generated and the prove

cycle will be reset to the start flag.

A deviation value in percentage for each parameter is calculated from the equation:

100

×

=

onditions

referencec

onditions

referencec

dition

currentcon

Deviation

Equation 5 Prove deviation calculation