SUMMIT 8800

STREAM CONFIGURATION

06

6 .2 .4 .2 Pulse Input API 5 .5 level B to E

Define the preference for serial or pulse input and the pulse input setting for API 5.5 level B to E

if used

Figure 79 Coriolis pulse input section for liquid and gas API 5 .5 Level B to E

Primary measurement

This option defines whether the serial link or the pulses will be used as a

primary measurement. The other will be used when the primary fails. Off

course only one of them can be used also: the other will not be connected..

Frequency method

Defines if the pulse corresponds with mass or volume. All Coriolis meters

measure mass, they can also calculate the volume. Though less accurate,

some users prefer this as their company standard

Impulse factor

The K-factor or impuls factor for the HF pulse. The total number of pulses

corresponding to one unit of flow.

Frequency input

API 5.5 level B to E.

Turbine minimum

frequency

Low cut-off frequency. This is frequency below which the flow will be

considered 0.

Blade ratio

Ratio between the two frequency inputs. For one input set to 0. For two

identical frequency inputs set to 1

Preset K Factor LF

The K-factor or impuls factor for the LF pulse. The total number of pulses

corresponding to one unit of flow. (Gas only).