SUMMIT 8800
STREAM CONFIGURATION
06
6 .2 .4 .2 Pulse Input API 5 .5 level B to E
Define the preference for serial or pulse input and the pulse input setting for API 5.5 level B to E
if used
Figure 79 Coriolis pulse input section for liquid and gas API 5 .5 Level B to E
Primary measurement
This option defines whether the serial link or the pulses will be used as a
primary measurement. The other will be used when the primary fails. Off
course only one of them can be used also: the other will not be connected..
Frequency method
Defines if the pulse corresponds with mass or volume. All Coriolis meters
measure mass, they can also calculate the volume. Though less accurate,
some users prefer this as their company standard
Impulse factor
The K-factor or impuls factor for the HF pulse. The total number of pulses
corresponding to one unit of flow.
Frequency input
API 5.5 level B to E.
Turbine minimum
frequency
Low cut-off frequency. This is frequency below which the flow will be
considered 0.
Blade ratio
Ratio between the two frequency inputs. For one input set to 0. For two
identical frequency inputs set to 1
Preset K Factor LF
The K-factor or impuls factor for the LF pulse. The total number of pulses
corresponding to one unit of flow. (Gas only).