KROHNE Summit 8800 Vol 2 User Manual
Page 162
162
www.krohne.com
08/2013 - MA SUMMIT 8800 Vol2 R02 en
SUMMIT 8800
SAMPLER
12
Figure 178 Sampler can flow limits
Sampler low and high flow
limit
Low and high flow limits to stop sampling
Automatically resume after
low and high flow
Restart sampling automatically after returning to normal
Sampler can
Also in case of volume measurement it is important to ensure that the sampler can is not over-
filled. For this reason there is an analog input for the measured can level “can input” in volume
% of full
This is compared to a can level calculated by the sampler, using the following parameters:
Figure 179 Sampler can calculated can level parameters
Sample can size
Can size in ml
Sample can maximum level
Alarm limit can level, in percentage of the ml
Sample can high level
Warning limit level, in percentage of the ml
Sample pulse duration
The time, in seconds, the output will be set to take one sample
Sample size
The quantity of product, in ml, of the sample being captured
Deviation alarms
It is important to verify the proper operation of the sampler. There could for instance be a faulty
valve which would not open or which is leaking. For this reason, there is the option to check for
deviation between the measured and calculated can level:
Figure 180 Sampler can calculated volume % full parameters