10

Figure 143 Prover options: general, proving points

Settings

Depending on the prover type, additional settings are needed as follows:

Figure 144 Prover options: general settings, uni-directional prover

Ball release time

Release command time to valve

Maximum loop time

Allowable time for sphere to travel a complete loop – chamber to chamber

Maximum chamber time

Time for sphere to travel past last detector switch to home chamber

Pump check delay time

Pump running signal check status wait delay

Figure 145 Prover options: general settings, bi-directional prover

Maximum loop time

Allowable time for sphere to travel a complete loop – chamber to chamber

Maximum chamber time

Time for sphere to travel past last detector switch to home chamber

Figure 146 Prover options: general settings, small volume prover

Maximum forward time

Piston travel time from start to end position

Maximum reverse time

Piston start position recovery time

Home detection

The use of a home detector switch – yes or no

Plenum deviation limit

Maximum deviations between calculated and measure plenumm pressure

Plenum deviation time

Deviation time allowable for limit