beautypg.com

KROHNE Summit 8800 Vol 2 User Manual

Page 161

background image

161

www.krohne.com

08/2013 - MA SUMMIT 8800 Vol2 R02 en

SUMMIT 8800

SAMPLER

12

Figure 176 Flow based sampler counter selection

Weighted sampler

Of course it is important to ensure that the sampler can is not overfilled. Here it is important to

know if the check is done on volume or mass.

In case of mass, the weight of both the possible cans is checked. Please note that this is a check

only against one or two analog inputs for the measured sampler can weight “can A and B input”

The parameters to check the cans are as follows:

Figure 177 Sampler can weighing

Sample measurement

Measurement based on volume or mass

Sample can mass unit

Selectable units for mass measurement – ibs or kg

Sample can weight deviation

Maximum deviation allowed between sampling and the weight of can A

and B

Same can maximum weight

Alarm limit for the weight of can A and B

Sample can high weight

Warning limit for the weight of can A and B

Sample can reset time

Change of can indication resetting all preset information (not

implemented yet)

Flow limits

It is important to get a representative sample in the can. If the flow is very low this might not be

the case. Simularly if the flow is too high, the flow to be counted in the error counters instead of

the normal counters, so the sample would not reflect the normal count. For this reason, limits

can be set outside which sampling may not occur: