KROHNE Summit 8800 Vol 2 User Manual

Page 171

background image

171

www.krohne.com

08/2013 - MA SUMMIT 8800 Vol2 R02 en

SUMMIT 8800

BATCHING

13

Figure 193 Batching parameters to be recalculated

For each listed parameter the user is present with a selection with the value used for the recal-

culation report:

Average

The average sum value for the given parameter

Keypad

A user defined fixed value

Start

The value determined at the start of the batch

End

The value determined at the end of the batch

Current

The current value being used at the time of report generation

This selection can be made including the keypad value for the following parameters:

CTSm

Meter expansion due to temperature

CPSm

Meter expansion due to pressure

MF

Meter factor

ρS

Standard density of the liquid

CSW

Correction sediment & water

KF

K-factor for volume (e.g. pulses/m3)

KF

K-factor for mass (e.g. pulses/kg)

USM correction

Corrections on the ultrasonic meter

Shrinkage

The shrinkage factor

YM

Linearisation curve of the meter

Fwv

Water Content Volume Factor

EF

Emission factor (CO2 calculation)

OF

Oxidation factor (CO2 calculation)

Hx

Heating value (inferior or superior)