beautypg.com

KROHNE Summit 8800 Vol 2 User Manual

Page 106

background image

106

www.krohne.com

08/2013 - MA SUMMIT 8800 Vol2 R02 en

SUMMIT 8800

STREAM CONFIGURATION

06

6 .8 .4 TAB measured

If measured or sensor is selected, in the TAB “measured” will be asked what to do with the Hart

units:

• Should an alarm given if there is a difference between the units specified and the units given

via Hart?

• If there is a difference should the units given via Hart replace the specified units?

Figure 101 Liquid and gas measurement selection

Also the pressure and temperature at the density point must be specified. Their selection is

almost identical to the stream pressure and temperature selection

6 .8 .5 TAB serial (liquid only)

If selected, the pressure and temperature at the density point must be specified in the TAB

“measured”and/or “serial”; this is not needed for gas or steam.

Figure 102 Liquid serial selection

Their selection is almost identical to the stream pressure and temperature selection. Please

note that in this case only 2 pressure and temperature sensors are possible but that also the

pressure and temperature at the meter can be used.

6 .8 .6 Line density table (includes TAB when liquid)

An additional icon for the line density table will appear in the left hand list:

Liquid

Gas

Steam

Liquid

Gas

Steam

This page is used to define the line density as a table of up to 10 pressures and 10 temperatures: