6 .13 .1 tab normal and extended – KROHNE Summit 8800 Vol 2 User Manual
Page 119
119
www.krohne.com
08/2013 - MA SUMMIT 8800 Vol2 R02 en
SUMMIT 8800
STREAM CONFIGURATION
06
Chromatograph A
The data from GC-A will be used
Chromatograph B
The data from GC-B will be used
Modbus
The data from a serial modbus interface will be used
Analog Input
The data from the analog input(s) will be used
Normalised
The value, as detemined by the selections, will be normalised
Used
The used value, as detemined by the selections, will be used
Gas component error determines how to handle the remaining components in case some of the
gas components are in error (alarm state). A choice can be made to use the other components
as they are or to use none of them.
As the gas components are in percentage of the total, all components together should be 100%.
If this is not the case the gas can be normalised: multiplied by a factor to make the total 100%.
The Normalization of Gas Data received can be selected from:
None
No normalisation needed: use the components as they are.
100%
Normalise all components to 100%
Raw GC Data
Normalise only the raw GC data and not the rest of the components
6 .13 .1 TAB Normal and extended
Sets Keypad and alarm values for gas components. The normal values refer to the 22 gas com-
ponents normally used in calculations. The extended give the other 23 components that can be
found in natural gas.
Figure 117 Normal Gas data
Per component enter the keypad value, the max and min alarm limits and the high and low
warning limits in % of the gas. If there is no need for an alarm or warning value, please leave
the associated field blank.