SUMMIT 8800

STREAM CONFIGURATION

06

Chromatograph A

The data from GC-A will be used

Chromatograph B

The data from GC-B will be used

Modbus

The data from a serial modbus interface will be used

Analog Input

The data from the analog input(s) will be used

Normalised

The value, as detemined by the selections, will be normalised

Used

The used value, as detemined by the selections, will be used

Gas component error determines how to handle the remaining components in case some of the

gas components are in error (alarm state). A choice can be made to use the other components

as they are or to use none of them.

As the gas components are in percentage of the total, all components together should be 100%.

If this is not the case the gas can be normalised: multiplied by a factor to make the total 100%.

The Normalization of Gas Data received can be selected from:

None

No normalisation needed: use the components as they are.

100%

Normalise all components to 100%

Raw GC Data

Normalise only the raw GC data and not the rest of the components

6 .13 .1 TAB Normal and extended

Sets Keypad and alarm values for gas components. The normal values refer to the 22 gas com-

ponents normally used in calculations. The extended give the other 23 components that can be

found in natural gas.

Figure 117 Normal Gas data

Per component enter the keypad value, the max and min alarm limits and the high and low

warning limits in % of the gas. If there is no need for an alarm or warning value, please leave

the associated field blank.