SUMMIT 8800
STREAM CONFIGURATION
06
Ultrasonic meter type
The specific meter to be used can be selected from a list. As meters are
generally designed for liquid, gas or stream, the list varies depending on
the medium. Also there is the option to use an analog input for flow.
Number of paths
The number of measurement paths (1 to 8) of the selected meter (see
meter specifications for details). Used when the meter communicates
serially. An alarm will be given if the meter is not configured for the same
number of paths.
Meter units
The engineering units in which the meter is measuring flow.
Maximum Counter
Increment
The maximum allowable increment for the flow data counter used for the
calculated volume. Used when the meter communicates serially. This
prevents a massive increment when, after a communication failure, the
meter resumes normal communication.
Flow offset
This is used to simulate flow during testing when no meter is available. In
theorie it could also be used to correct a fixed mis-match of the flow.
Meter specific data
For certain meters some parameters must be set. As these parameters
are meter specific, please consult the meter manufacturer’s operating
instructions for further guidance. Parameters for the following meters are
available:
- Flowsic 600
Use or ignore the security alarm, provided by Modbus or a digital output
- Senior Sonic
Dimensions and material of the meter and calculation of dynamic Viscosity
- Instromet QSonic
UFM efficiency settings used for reliability and diagnostic information.
- KROHNE UFM 3030
Select if the measured flow rate, velocity of sound and totals should be
used.
6 .2 .2 .2 Pulse Input API 5 .5 level B to E
Define the preference for serial or pulse input and the pulse input setting for API 5.5 level B to E
if used
Figure 69 Ultrasonic pulse input section for liquid and gas API 5 .5 Level B to E