SUMMIT 8800

STREAM CONFIGURATION

06

Enter as a percentage the nozzle pressure loss.

ISA 193 2 nozzle

Typically used for high velocity, set by ISO 5167 to determine the flow of fluid.

Enter as a percentage the nozzle pressure loss.

Long radius

nozzle

A variation of the ISA 1932 nozzle, with a convergent section as the ISA 1932

nozzle and divergent section as a classical venture. Enter as a percentage the

nozzle pressure loss.

6 .2 .3 .1 .4 Cone meter

Cone

Select the V-cone meter to be used.
NUFLO
McCrometer
McCrometer wafer

Keypad CoD

A user defined value for the coefficient of discharge

6 .2 .3 .2 Meter information

For each meter information to identify the meter can be enetered. This can be usefull as identi-

fier text on the screen or to send to a supervisory system in a system.

Figure 74 Define meter information

6 .2 .3 .3 Coefficient of Discharge Table

Liquid

Gas

Steam

This section applies to the classical venture only. Here is an option to select a CoD table. The

ratio of the actual discharge to the theoretical discharge:

CoD = Actual discharge / Theoretical discharge.

When selected, this page is used to set the values of CoD to be used in that table. The CoD can

be given for 1 to 10 values of Pressure and Temperature. Linear interpolation is carried out

between the table values.