SUMMIT 8800

STREAM CONFIGURATION

06

Correction points

The size of the table can be selected from 2 to 30 points or as none: no

correction.

Correction type

Linear, where the correction is applied through the operating range

by linear interpolation between the two nearest data points and linear

extrapolation for low and high flow.

MID, where the correction is limited under Low and high flow conditions

as required by the MID approval. There is no extrapolation done for

these conditions.

%Qmax-%Error

A data set of a flow rate (as a percentage of Qmax) and the associated

error (%). The Qmax refers to the meter at hand not the reference

meter. Always start with the lowest flow rate and each following must

be higher than the previous.

%Qmax is in % of maximum flow (Q max.) of the meter and can range

from – Qmax to + Qmax to allow for different linearity in both flow

direction

%Er.rd is the % error of reading. If the output from the meter is less

than the actual flow through the meter then the error is entered as a

negative value.

Calculate

A calculation button allows the user to simply check the correction

that would be applied at any flow rate entered. In case the flow for

the reference meter is given, the Qmax for the meter at hand can be

calculated and v.v.:

Figure 88 Gas or steam flow rate calculations

Linear correction is only applied if the meter produces at least 10 pulses per second at Qmin.