APPENDIX 2: GAS CALCULATIONS

17

17. Appendix 3: Gas calculations

17 .1 Perform meter body correction

If the operating temperature and pressure is different from the conditions at high pressure cali-

bration, the cross sectional area of the flow meter will change slightly. The following correction

can be applied to the raw flow rate to compensate for this. The basic configuration has no meter

body correction applied.

Meter body correction calculations

The system corrects the volume flow for the impact of pressure and temperature on the geom-

etry of the flanged meter body according to ISO standard 17089-1.

Simplified direct single-stage calculation is used which estimates the effect for pressure and

temperature:

1) Body temperature effect

(

)

T

=

Q

Q

ature

bodytemper

+





*

*

3

1

0

1

α

Equation 21 Meter body temperature effect

Where:

Q1/Q0

Flow correction factor for the meter output [-]

α

Thermal expansion coefficient [K-1]

ΔT

Temperature difference [K]

With:

cal

op

T

T

=

T

Equation 22 Delta meter body temperature

Where:

Top

Field operation temperature [K]

Tcal

Calibration temperature [K]

2) Body pressure expansion









+

+

+





E

P

R

a

R

a

=

Q

Q

re

bodypressu

*

*

4

1

2

2

2

2

max

0

1

σ

Equation 23 Meter body expansion

Where:

Q1/Q0

Correction factor for the meter output [-]

R

Inside pipe radius [m]

α

Outside diameter of the meter body [m]

σ

Poison’s ratio [-]

ΔP

Pressure difference [Pa]

E

Modulus of elasticity [Pa]