beautypg.com

6 .13 gas data – KROHNE Summit 8800 Vol 2 User Manual

Page 118

background image

118

www.krohne.com

08/2013 - MA SUMMIT 8800 Vol2 R02 en

SUMMIT 8800

STREAM CONFIGURATION

06

6 .13 Gas Data

Liquid

Gas

Steam

For liquids, gas data will only be needed if LNG or LPG is used, so if the line density is calcu-

lated.

This section defines the selection process to arrive to “in use” gas components. There are 4 se-

lections in order of priority with each 9 choices. The SUMMIT will start to check the first selec-

tion. If the value is valid (not in alarm) then this selection will be choosen. If not, then the second

selection will bechecked etc. The last selection is always the keypad value as defined earlier,

even if all other choices are none.

The SUMMIT goes through the selection on a per component base. So if a GC would be the first

choice and the GC only provides the data to C6+, then the other components will be taken from

the next choices. This means that also the moisture can be added as a selection though a GC

will not provide this.

The 9 choices provided for the gas data are:

Keypad

The fixed value enetered under TAB’s Normal and extended are used.

Chromatograph A

The data from GC-A will be used

Chromatograph B

The data from GC-B will be used

Modbus

The data from a serial modbus interface will be used

Analog Input

The data from the analog input(s) will be used

Last Good Value

The last good value before the selection went into error will be used

Last Hour Average

The value from the last hour will be used

Last Day Average

The value from the last day will be used

Calculated H2O value

The value for moisture will be used to calculate the total componenent

Figure 116 Gas data selection

A separate choice can be made to determine the source for the average gas composition: