KROHNE Summit 8800 Vol 2 User Manual
Page 112
112
www.krohne.com
08/2013 - MA SUMMIT 8800 Vol2 R02 en
SUMMIT 8800
STREAM CONFIGURATION
06
6 .10 .2 Relative density / Specific gravity
Liquid
Gas
Steam
This section defines the selection process to arrive to an “in use” value for the relative density.
There are 3 selections in order of priority with each 10 choices. The SUMMIT will start to check
the first selection. If the value is valid (not in alarm) then this selection will be chosen. If not,
then the second selection will be checked etc. The last selection is always the keypad value as
defined later, even if all other choices are none.
Figure 111 Relative density options
The 10 choices provided for the relative density selection are:
Keypad
The fixed value entered will be used
Chromatograph A
The value from GC-A will be used
Chromatograph B
The value from GC-B will be used
Modbus
The value from a serial modbus interface will be used
Sensor
The value from the analog or Hart input will be used
Extra inputs appear for the calibration constants of this sensor.
ISO 6976
The value is calculated using the ISO 6976 standard, from the gas data
This requires an extra input of the reference temperature of the rd.
Last Good Value
The last good value before the selection went into error will be used
Last Hour Average
The value from the last hour will be used
Last Day Average
The value from the last day will be used
GPA2172
The value is calculated using the GPA2172 standard, from the gas data
An extra page will appear for specific data related to the parameter.