beautypg.com

Teledyne LeCroy Summit T3-16 PCIe Multi-lane Protocol Analyzer User Manual User Manual

Page 115

background image

    

Summit T3‐16 PCI Express Multi‐Lane Protocol Analyzer User Manual

103

Probe Settings

 Teledyne  LeCroy

To use the external reference clock or for quick calibration click the appropriate 
checkboxes to select them. If checked, calibration routine will only scan CTLE, if not 
checked it will include DFE with the CTLE scan which will take longer, approximately 9 
minutes.

Figure 5.53: Calibrate Dialog

The calculated CTLE values are automatically displayed in the relevant lane field. Click 
Save to save the settings or click Save As Default to save the settings as the default. Click 
Load to save the settings as a .rec file as a PE Tracer Recording option to be used at a later 
time. Click Reset to clear the settings.

The calculated CTLE values are automatically displayed in the relevant lane fields as 
shown in Figure 5.54. Click Save to save the settings or click Save As Default to save the 
settings as the default. Click Load to save the settings as a .rec file as a PE Tracer 
Recording option to be used at a later time. Click Reset to clear the settings (see 

Figure 5.54 on page 104

).