beautypg.com

25 filter out, 26 probe settings – Teledyne LeCroy Summit T3-16 PCIe Multi-lane Protocol Analyzer User Manual User Manual

Page 113

background image

    

Summit T3‐16 PCI Express Multi‐Lane Protocol Analyzer User Manual

101

Filter Out

 Teledyne  LeCroy

5.25 Filter

Out

A filter causes the Analyzer to filter out specified events from the recording. If events are 
filtered out of the recording, they are excluded from the Analyzer's buffer and not simply 
hidden from the CATC Trace. The purpose of filtering is to preserve recording memory so 
you conduct longer recording sessions and exclude events that do not interest you.

To enable or disable filtering, place an event button into either the Global State cell or 
the Sequence cell, then right‐click the button and choose an Action (see previous topic). 

From the sub‐menu, select Filter Out.

Figure 5.51: Filter Out Option

Note:

The Filter Out option is only available in the DLLP, TLP, and Ordered Sets TS1 and TS2.

5.26 Probe

Settings

The Summit T3 analyzer can be optimized for specific DUT signal characteristics. The auto 
calibration feature in the Probe Settings allows for better signal recovery. There are 16, 0 
db to 15 db, internal equalization modes in the analyzer hardware that boost the high 
frequency content of the captured signal. This equalization does not impact the signal to 
the DUTs, but helps in ‘opening the eye’ of the captured signal, so that the internal logic 
can capture PCIe frames. The system is designed to allow you to manually set the 
equalization to one of the sixteen modes for the CTLE values and one of the eight modes 
for the DFE values or choose to run an automated algorithm, internal to the analyzer. This 
automated calibration sets the optimized values for each of the PCIe captured lanes, 
provided that the signal at the analyzer input is stable.

Each lane can be configured to use specific equalization settings to match signal 
characteristics of the lane and improve the analyzer’s ability to capture error free traces. 
For 2.5/5.0 GT/s speeds only CTLE setting is required to achieve good captures. For 8.0 GT/
s you can adjust CTLE, DFE and DC Gain parameters.

Select Recording Options > Probe Settings from the Setup menu to display Probe 
Settings.