LabMaster 10 Zi Oscilloscopes

80

LM10Zi-GSM-E Rev A

instances when a Num Segments miscount occurs for some
reason and the oscilloscope waits indefinitely for an
unforthcoming segment. During that time, no oscilloscope
functions are accessible. By means of a timeout value, however,
the acquisition will be completed, the waveform displayed, and
control of the oscilloscope returned to the user after the timeout
has elapsed.

6.

Under Display Settings, touch inside the Display mode control,
and select a sequence mode display from the pop-up menu.

NOTE: Once a single acquisition has started, you can interrupt the
acquisition at any time by pressing the S

INGLE

front panel button

a second time or by pressing the S

TOP

front panel button. In this

case, the segments already acquired will be retained in memory.

7.

Touch the

S

INGLE

trigger front panel button.