Use the Auxiliary Output to send the following signals to other
instruments:

Trigger Enabled - This function can be used as a gating

function to trigger another instrument when the oscilloscope is
ready.

Trigger Out - This function can be used to trigger an

external oscilloscope.

Pass/Fail - Allows you to set a pulse duration (which

varies based on your specific model, refer to datasheet
specifications at www.teledynelecroy.com for details); generates
a pulse when pass/fail testing is active and conditions are met.

Fast Edge - Some models provide this signal output used

to deskew multiple channels. The Fast Edge connection varies
based on model and is either SMA or BNC.

C

ALIBRATION

O

UTPUT

S

IGNALS

A calibration signal can be output from the Probe Calibration Hook on
the front of the oscilloscope. The following Calibration Output Signals
are available for use:

Square Wave - You can set a Frequency value (varied

based on your specific model, refer to datasheet specifications at
www.teledynelecroy.com for details).

DC Level - This is a reference level.

Off - This function turns off the Auxiliary Output signal.